5.0.2
Advanced Entry Calibration
 
 Advanced Entry Calibration is the alternate method for special-case media
types. The procedure has two parts:
 
 1)  Using different sampling algorithms, displayed sensor readings have to be
listed for each label and TOF value.
 
 2)  From this data, the best algorithm is chosen; then those values are manually
entered for the ‘Paper’, ‘Gap’ (or ‘Mark’), and ‘Empty’ variables.
 
+
Advanced Entry Calibration will override all previous calibration
settings; this method should be used only when Standard Calibration
has failed.
 
 To perform an Advanced Entry Calibration:
  
 Step
 Operator Action
 Displayed Message
 Comment
 
 1
 
 
 (Media should be loaded
and the appropriate
Sensor Type selected.)
 
 Turn ‘On’ the printer.
 
 
 
 
 
 
 Wait until the
printer initializes
(about six
seconds).
 
 
 2
 
 
 Press the 
 
key.
 
 Raise the printhead
assembly.
 
 
 
 
 
  
 
 You are entering
Menu Mode.
 
 
 
 3
 
 Press the                key to
enter the Media Settings
menu.
 
 
 
 
 
  
 
 See Section 4.1.2
for menu layout.
 
 4
 
 Press the                key
and scroll to ‘Sensor
Calibration’.
 
 
 
 
 
 
 Press the
 key to abort this
procedure.
 
 
 5
 
 
 Press the
                       
 
key.
 
 
 
 
 
 
 
 You are entering
the calibration
submenu.