0611 
RM500SL User’s Guide Version 2.8  
Page 47 
9.3 
Multicurve procedure 
FastFacts 9.3: Multicurve procedure 
9.4 
Multicurve results 
FastFacts 9.4: Multicurve results 
Multicurve generates 1 to 4 output or gain response curves using either pink noise 
or swept pure tones at input levels ranging from 40 – 90 dB SPL and 0 dB (off). 
1)  Press <Tests>, then highlight and <PICK> [Multicurve]. 
2)  Refer to 
7.11: Coupling the hearing instrument
7.12: Positioning the hearing 
instrument
 and 
2.5: Battery pill use and care
 to prepare the instrument for 
testing. Turn the instrument ON. 
3)  Highlight and <PICK> button 1, 2, 3 or 4 in the Test column. 
4)  Highlight and <PICK> the Stimulus window, then highlight and <PICK> either 
[Pink noise] or [Swept] from the drop-down list. 
5)  Highlight and <PICK> the Level window for the selected Test, then highlight 
and <PICK> a stimulus level from the drop-down list. 
6)  Press <Continue> to run a sweep or to capture a pink noise curve; press 
<Cancel> to abort the test. 
7)  To measure another response curve, highlight and <PICK> another test button; 
to overwrite a curve, highlight and <PICK> its test button. 
To change scale from dB SPL to dB Gain, highlight & <PICK> the Scale window. 
1)  Output curves for the noise signal are in 1/12
th
 octave bands. They will be 
approximately 18 dB lower than pure tone curves for a linear instrument. 
2)  Gain curves for a linear instrument will be the same for both swept and pink 
noise test signals. 
3)  For compression instruments, both gain and output curves may be expected to 
differ for the two test signals. 
4)  See 
7: Hearing Instrument Tests Setup
 to change screen appearance.