9.7 
Manual control procedure 
FastFacts 9.7: Manual control procedure 
9.8 
Sound level meter using manual control 
FastFacts 9.8: Sound level meter using manual control 
Manual control displays the reference microphone SPL, the coupler microphone 
SPL, their difference in dB (gain) and coupler microphone signal distortion at any 
selected 1/12
th
 octave frequency and at a selected input level from 40 to 90 dB (in 
5 dB steps) and at 0 dB (off). 
1)  Press <Tests>, then highlight and 
<PICK> [Manual control]. 
2)  Highlight and <PICK> the Level 
window, then highlight and <PICK> the 
desired stimulus level from the drop-
down list. 
3)  Highlight and <PICK> the Frequency 
window, then highlight and <PICK> the 
desired frequency from the drop-down 
list. (Distortion measurement data is 
available only from 200-4000 Hz.) 
4)  Highlight and <PICK> [Start test] to 
start the test. [Start test] changes to 
[Stop Test]. To stop the measurement, 
highlight and <PICK> [Stop test]. You 
may change Level and Frequency 
without stopping the test. 
When set for a stimulus level of 0 dB (off), the Manual control mode allows the 
RM500SL to be used as a sound level meter. Sound at the HIT reference 
microphone and the coupler microphone may be analyzed by 1/12
th
 octave, A-
weighted or C-weighted filters. A and C-weighted measurements are limited to 
frequencies above 100 Hz and the averaging time and update rate are 384 ms. 
1)  Press <Tests>, then highlight and <PICK> [Manual control]. 
2)  Highlight and <PICK> the Level window, then highlight and <PICK> 0 dB 
stimulus level from the drop-down list. 
3)  Highlight and <PICK> the Filter window, then highlight and <PICK> 1/12
th
 
octave, A-weighted, or C-weighted from the drop-down list. 
4)  For 1/12
th
 octave filter, highlight and <PICK> the Frequency window, then 
highlight and <PICK> the desired frequency from the drop-down list. 
5)  Highlight and <PICK> [Start test] to start the test. [Start test] changes to [Stop 
Test]. To stop the measurement, highlight and <PICK> [Stop test]. You may 
change Level and Frequency without stopping the test.  
NOTE: A microphone extension cable (VA-130) is available from Audioscan. 
Standard audio extension cables should not be used. They will substantially 
increase noise levels.