HP (Hewlett-Packard) RM500SL Manuel D’Utilisation

Page de 112
0611 
RM500SL User’s Guide Version 2.8  
Page 55 
11  REM Screen Setup 
11.1  Section overview 
FastFacts 11.1: REM screen setup 
11.2  REM right or left ear display 
FastFacts 11.2: REM right or left ear display 
11.3  Graph, table or 2cc target format 
FastFacts 11.3: Graph, table or 2cc target format 
Most REM test screens may be viewed in graphic or tabular format. 
v  In RECD, highlight and <PICK> [Format] to toggle between [Graph] and 
[Table] views. 
v  In Speechmap and Insertion gain, highlight and <PICK> [Format], then 
highlight and <PICK> [Graph], [Table] or [2cc targets] from the drop-down list. 
v  In [2cc targets] format, reserve gain and fitting formula may be changed. 
Highlight & <PICK> the appropriate window, then highlight & <PICK> the 
reserve gain or fitting formula from the drop-down list. 
In Speechmap, if DSL or NAL-NL1 has been selected as a target method, then 
2cc coupler targets for user gain, full-on gain and OSPL90 are provided. The full-
on gain is just the user gain plus the reserve gain. If no target method has been 
selected, only 2cc OSPL90 values are indicated. Target gains are for speech-
like signals at a 70 dB input SPL; OSPL90 targets are for 90 dB pure tones.
 
Real-ear targets are converted to 2cc coupler values by using age-appropriate 
average or measured (if available) RECD values and the appropriate microphone 
location effects for the desired style of hearing instrument.  
The gain values given assume the hearing instrument is linear for a 70 dB speech 
input. This assumption is necessary to allow the microphone location effects to be 
incorporated. 
For BTE and body aids, the 2cc coupler values provided assume a fully-occluding 
earmold with #13 tubing extending to the medial tip of the earmold. If a different 
type of mold is used, alterations to these values should be made to account for the 
different acoustic characteristics. For CIC instruments, the 2cc values take into 
account that the instrument, when deeply inserted into the ear canal, will produce 
a greater SPL in the ear than that measured in the ITE coupler. 
The RM500SL can retain and display REM test results for two ears.  To switch 
between left and right test results, press the <Left/Right> key. 
To accommodate your working preferences, the appearance of REM screens may 
be changed in several ways.  
This section covers the following topics: 
11.2: REM right or left ear display 
11.4: SPL or HL scale 
11.3: Graph, table or 2cc target format 
11.5: Hide or show REM curves