11.4  SPL or HL scale 
FastFacts 11.4: SPL or HL scale 
11.5  Hide or show REM curves 
FastFacts 11.5: Hide or show REM curves 
12   REM Manual Control 
12.1  Section overview 
FastFacts 12.1: REM Manual Control 
In Speechmap and Insertion gain tests you may view results in either dB HL or dB 
SPL. 
Insertion gain tests view the hearing aid as improving the sound-field threshold. 
In Insertion gain, the HL scale presents aided results as improved sound-field 
thresholds while the speech region remains fixed. 
Speechmap tests view the hearing aid as increasing the ear canal SPL. Speech 
gets louder but the threshold remains fixed. On the HL scale in Speechmap, 
normal ear canal threshold becomes the 0 dB HL line and all data is replotted 
using this reference. A physical interpretation of the data is difficult and the use of 
HL in Speechmap is not recommended. All instructions given in this guide 
assume the SPL Speechmap scale. 
To change the scale: 
Highlight and <PICK> [Scale] to toggle between a dB SPL and dB HL scale. 
On screens capable of showing more than one curve, you may show or hide test 
curves. Hiding a curve does not erase it. To erase curves use the <Session> key. 
To hide or show test curve(s): 
1)  Highlight, then <PICK> Curve [Hide/Show]. 
2)  Highlight, then <PICK> the desired test curve to toggle between Hide or Show. 
The curve will toggle on or off when selected. 
3)  Press <Continue> to return to the test screen. 
Manual control displays the reference microphone SPL, the probe microphone 
SPL, their difference in dB (gain) and probe microphone signal distortion at any 
selected 1/12
th
 octave frequency and at a selected input level from 40 to 90 dB (in 
5 dB steps) and at 0 dB (off). With the stimulus turned off, the probe module 
becomes a dual channel sound level meter with a choice of 1/12
th
, A and C filters 
in both channels. 
This section covers the following topics: 
12.2: Manual control measurements 
12.3: Sound level meter using REM probe