RM500SL User’s Guide Version 2.8  
Page 63 
15.5  RECD results 
FastFacts 15.5: RECD results 
The RECD measurement screen shows the response of the RECD transducer in 
the real ear (pink) and in the 2cc coupler (green). The difference between these 
curves is the RECD (cyan). An age-appropriate average RECD is shown for 
reference (dotted). A negative RECD below about 1 kHz may indicate a poor seal 
between the foam tip and the ear canal. A negative RECD above about 3 kHz may 
indicate the foam tip is blocking the probe tube. See 
15.2: RECD facts
15.3: 
RECD coupler response
15.4: RECD real-ear response