15.3  RECD coupler response 
FastFacts 15.3: RECD coupler response 
Prepare the system for HIT tests. See 
2.3: Microphone connection
7.9: 
Calibration of HIT reference microphone
. Connect the optional RECD transducer 
to the RECD jack to the left of the test chamber. 
1)  Press <Tests> then highlight and 
<PICK> [RECD] from the REM 
tests list. 
2)  If a valid coupler response exists, 
go to 
15.4:RECD real-ear 
response. 
3)  If no coupler response exists or the 
measurement interval has been 
exceeded, connect the RECD 
transducer to the BTE coupler. 
4)  Highlight and <PICK> [Measure coupler] and follow the setup instructions. 
5)  Press <Continue> to generate the coupler response curve (green). When this 
curve is stable, press <Continue> to save it. 
6)  Highlight and <PICK> [Daily] or [Weekly] to toggle the measurement interval for 
the coupler response. 
For comparison purposes, an age-related average RECD is shown as a dotted 
curve on the screen. If you have accessed RECD from the REM tests list, you may 
highlight and <PICK> the Age box to change the age for this reference curve.