Leica CM3050 S – Cryostat
6.
Troubleshooting, applications tips
Problem
Sections tear.
Inconsistent or insufficient
specimen feed.
Specimen disc cannot be
removed.
Cryostat inoperational.
Sectioning motor inoperational.
No or insufficient refrigerating
performance.
Causes
- Temperature too low for the
tissue to be sectioned.
- Knife blunt, dirty, dusty, frosted
or rusty.
- Upper edge of anti-roll plate
damaged.
- Hard particles in the tissue.
- Rear face of knife dirty.
- Microtome was not completely
dry when refrigeration was
switched on; ice build-up inside
the micrometer mechanism.
- Microtome defective.
- Specimen disc stuck to quick-
freeze shelf or specimen head
due to moisture on underside
of specimen disc.
- Mains plug not properly con-
nected.
- Automatic cut-out has been
triggered.
- Footswitch not connected.
- Automatic cut-out of sectioning
motor has been                    trig-
gered.
- Motor defective.
- Stopper not placed properly
into drain in bottom of cryo-
chamber.
- Compressor defective.
- Leak in cooling system.
Remedies
- Select higher temperature and
wait for the new temperature
to be reached.
- Eliminate cause (--> see details
on individual problems).
- Exchange anti-roll plate.
- If application permits, section
at lower plane.
- Clean.
 - Remove microtome from cham-
ber and dry - possibly with hair-
dryer (hairdryer only to be used
outside the cryochamber!) or at
ambient temperature.
- Call Technical Service.
- Apply strong alcohol to contact
point or heat the specimen
head.
- Check mains plug connection.
- Reconnect automatic cut-out
(switch instrument back on).
- Connect footswitch.
- Reconnect automatic cut-out
(push inwards until it locks in
place).
- Call Technical Service.
- Close drain with stopper.
- Call Technical Service.
- Call Technical Service.