Leica DM4000B Manuel D’Utilisation

Page de 69
20
5. Assembly
5.5 Light Sources for the Transmitted Light Axis
Caution:
Be sure that the lamp housing is discon-
nected from the power supply. Unplug the
power plug and the power supply during as-
sembly.
107/2 Lamp Housing
This lamp housing is used with a 12V 100W halo-
gen lamp, which is already mounted.
In case the lamp has to be removed:
• Remove the fastener screw on the housing
(Fig. 12).
• Remove the housing by pulling it upwards.
• Remove the lamp
• Insert the new 12V 100W lamp (13.1) with dust
cover straight into the socket until it stops. Be
sure that the lamp is inserted straight.
• Remove the lamp’s dust cover.
Caution:
Do not remove the lamp’s dust cover until
you have installed the lamp. Avoid
fingerprints on the lamp.
• Replace the housing and fasten it in place us-
ing the fastening screw.
• Place the lamp housing in the transmitted light
lamp housing receptacle (14.2) and fasten it
with the clamping screw on the side.
• Connect the lamp housing to the power
supplyfor transmitted light (symbol:  ) (14.3).
Fig. 12
Lamp housing 107/2
Releasing the
fastening screw
1
2
Fig. 13
Lamp housing 107/2,
opened
Mount with
halogen lamp
Collector
Fig. 14   Rear side of stand
Incident light lamp housing receptacle
Transmitted light lamp housing receptacle
12 V 100 W connection for transmitted light (symbol:  )
12 V 100 W connection for incident light (symbol:      )
1
2
3 4