Leica DM4000B Manuel D’Utilisation

Page de 69
21
5. Assembly
5.6 Light Sources for the Incident Light Axis
Caution:
During assembly, always unplug the power
supply unit of the 106 z lamp housing from its
socket.
Never touch the glass parts of the burner
with bare hands.
Never look directly into the beam path (blind-
ing hazard).
During assembly work on xenon burners, al-
ways wear the supplied protective gloves and
face protection (Fig. 15) (risk of explosion).
106 z lamp housing
This lamp housing is used with a 12V 100W halo-
gen lamp or various gas discharge lamps.
Inserting the 12V 100W halogen lamp into the
106 z lamp housing
• Unscrew the fastening screws of the cover
and lift up the cover (16.1).
• Unscrew the fastening screws of the lamp
mount (16.8) and pull out the mount (Fig. 17).
Fig. 15
Protective gloves and mask
Fig. 16     106 z lamp housing (on the side, open)
1
Cover raised
2
Collector
3
12 V 100 W lamp or
gas discharge lamp in mount
4
Reflector (mirror)
5, 6, 7 Adjusting screw for x-y reflector
8
Fastening screw for lamp mount
9
Socket for contact plug
1
2
3
8
9
8
4
5
6
7