FluoCombi III™ – Adjusting the Parfocality
Setting up the objective nosepiece
 
The  following  configuration  only  has  to 
be  carried  out  once.  It  ensures  that  the 
parfocality remains intact, and thus the speci-
men  does  not  need  to  be  brought  into  focus 
again.
1.  Place a specimen under the microscope.
2.  Rotate the stereo planachromatic objective 
under the beam path.
3.  Reduce  the  view  so  that  you  can  see  the 
specimen in the overview.
4.  Center the specimen under the beam path.
5.  Adjust  the  magnification  to  the  highest 
level and focus on a flat specimen.
6.  Rotate  the  5×  planachromat  under  the 
beam path.
7.  Unscrew  the  clamping  screw  on  the 
 objective.
8.  Reduce  the  view  so  that  you  can  see  the 
specimen in the overview.