Philips MP40/50 Manuale Utente

Pagina di 414
18 Monitoring EEG
EEG Monitoring Setup
208
EEG Monitoring Setup
1
Plug the trunk cable into the EEG module in the Flexible Module Server.
2
Prepare the patient’s skin prior to placing the electrodes. Good electrode-to-skin contact is 
important for a good EEG signal, as the skin is a poor conductor of electricity.
– Shave hair from sites, if necessary.
– Wash sites thoroughly with soap and water. We do not recommend using ether or pure alcohol, 
because this dries the skin and increases the resistance.
– Use a skin preparation paste to remove skin cells and oil before placing the electrodes. 
3
Select the desired electrode montage in the 
Setup EEG
 menu or in the 
EEG Impedance/
Montage
 window. 
4
Attach the reference electrode first. 
5
Place the electrodes on the patient’s head according to the selected montage. Use electrode gel if 
you are not using pre-gelled electrodes. Remember to select a site where the signal will not be 
interfered with by muscle artifacts. 
6
Connect the electrode connector end to the trunk cable.
7
Check the electrode-to-skin impedance in the 
EEG Impedance/Montage
 window.
8
For good signal quality, keep all lead wires together and away from other electric devices and 
metallic bodies.
Using the EEG Impedance/Montage Window 
To open the window, in the 
Setup EEG
 menu, select 
Show Montage
, or select the 
EEG Montage
 SmartKey.
The window may be configured to look slightly different on your monitor. 
ImpedanceLimit:5kOhm
Montage A
EEG 2: F4-P4
EEG 1: F3-P3
EEG Impedance/Montage
+
+
-
-
EEG 1
EEG 2
1
1
1
1
kOhm
Wiring and impedance values for the 
selected montage
Electrode locations on the patient’s head. The symbols 
represent the electrode-to-skin impedance.