Philips MP40/50 Manuale Utente

Pagina di 414
Using the EEG Impedance/Montage Window
18 Monitoring EEG
209
Choosing an EEG Electrode Montage
1
To activate one of the five pre-configured electrode montages, select the arrow beside the label in 
the 
EEG Impedance/Montage
 window and choose a montage from the list. 
2
Attach the electrodes as illustrated in the 
EEG Impedance/Montage
 window.
The five default electrode montage configurations can be modified and renamed in Configuration 
Mode.
The electrode locations are labeled according to the international 10-20 electrode placement system. 
Changing the Impedance Limit
The impedance limit can be set for all electrodes simultaneously in the 
Setup EEG
 menu, or in the 
EEG Impedance/Montage
 window using the pop-up keys. If the limit is exceeded during 
monitoring, an INOP will appear and the graphic impedance indicator will change.
To change the impedance limit, either 
use the pop-up keys that appear with the 
EEG Impedance /Montage
 window, or 
in the 
Setup EEG
 menu, select 
Impedance Limit
 to call up a list of selections between 1 and 
30 kOhm, then select the required limit from this list.
Montage Name
EEG1+
EEG1- Label1
EEG2+ EEG2-
Label2
Mont.A
Fp1
T3
Fp1-T3
Fp2
T4
Fp2-T4
Mont.B
O1
T3
O1-T3
O2
T4
O2-T4
Mont.C
F3
C3
F3-C3
F4
C4
F4-C4
Mont.D
C3
P3
C3-P3
C4
P4
C4-P4
Mont.E
Fp1
T5
Fp1-T5
Fp2
T6
Fp2-T6