SECTION 7—CYLINDER FILL TIMES
24
SECTION 7—CYLINDER FILL TIMES
NOTE: All filling times are approximate and may vary on 
environmental conditions.
*
Troubleshooting
If Conserving Cylinder fails to function, consult the 
Troubleshooting guide. If problem cannot be corrected, 
consult your Dealer/Home Care equipment provider.
Cylinder 
Filling 
Times
Concentrator Flow Rate to Patient:
(Platinum 5 up to 3 L/min. and Platinum 10 up to 5 
L/min)
HF2PCM2KIT
NORMAL 
25 min
MAXIMUM
30 min
PROBLEM
POSSIBLE CAUSE
SOLUTION
No flow.
1. Regulator in “Off” 
position.
1. Set to prescribed 
setting.
2. Cylinder empty.
2. Replace/fill cylinder.
3. Dial not set at 
increment.
3. Set dial so indictor 
points to a setting.
4. Conserver not 
sensing breath.
4. Check position of 
cannula in nose. Do not 
breathe through mouth.
5. Check to see if the 
cannula is kinked.