SECTION 7—CYLINDER FILL TIMES
24
SECTION 7—CYLINDER FILL TIMES
NOTE: All filling times are approximate and may vary on
environmental conditions.
*
Troubleshooting
If Conserving Cylinder fails to function, consult the
Troubleshooting guide. If problem cannot be corrected,
consult your Dealer/Home Care equipment provider.
Cylinder
Filling
Times
Concentrator Flow Rate to Patient:
(Platinum 5 up to 3 L/min. and Platinum 10 up to 5
L/min)
HF2PCM2KIT
NORMAL
25 min
MAXIMUM
30 min
PROBLEM
POSSIBLE CAUSE
SOLUTION
No flow.
1. Regulator in “Off”
position.
1. Set to prescribed
setting.
2. Cylinder empty.
2. Replace/fill cylinder.
3. Dial not set at
increment.
3. Set dial so indictor
points to a setting.
4. Conserver not
sensing breath.
4. Check position of
cannula in nose. Do not
breathe through mouth.
5. Check to see if the
5. Check to see if the
cannula is kinked.