Invacare IRC5PAW Manuale Utente

Pagina di 32
SECTION 8—TROUBLESHOOTING GUIDE
27
Perfecto
2
™Series
SECTION 8—TROUBLESHOOTING 
GUIDE
SYMPTOM:
PROBABLE CAUSE:
SOLUTION:
Alarm:
Main Power Loss:
Short beeps, long 
pause
1. Power cord not 
plugged in.
1. Insert plug into outlet.
Concentrator not 
operating,
power switch On.
2. No power at outlet.
2. Inspect house circuit 
breakers or fuses. If problem 
recurs, use a different outlet.
Beep......Beep......
3. Tripped circuit 
breaker.
3. Push/reset circuit breaker. 
If problem recurs, call Service 
Provider. 
Alarm:
Continuous
Concentrator not 
operating, Power 
Switch On. 
Beeeeeeeep....
System Failure:
1. Unit overheating due 
to blocked air intake.
2. Insufficient power at 
outlet.
3. Internal repairs 
required.
1a. Remove and clean cabinet 
filters.
1b. Move oxygen 
concentrator at least three 
inches away from walls, 
draperies or furniture.
2. DO NOT use extension 
cords. Move to another 
electrical outlet or circuit.
3. Call Service Provider.