Справочник Пользователя для Invacare IRC5PAW
SECTION 8—TROUBLESHOOTING GUIDE
27
Perfecto
2
™Series
SECTION 8—TROUBLESHOOTING
GUIDE
GUIDE
SYMPTOM:
PROBABLE CAUSE:
SOLUTION:
Alarm:
Main Power Loss:
Short beeps, long
pause
pause
1. Power cord not
plugged in.
plugged in.
1. Insert plug into outlet.
Concentrator not
operating,
power switch On.
operating,
power switch On.
2. No power at outlet.
2. Inspect house circuit
breakers or fuses. If problem
recurs, use a different outlet.
breakers or fuses. If problem
recurs, use a different outlet.
Beep......Beep......
3. Tripped circuit
breaker.
breaker.
3. Push/reset circuit breaker.
If problem recurs, call Service
Provider.
If problem recurs, call Service
Provider.
Alarm:
Continuous
Continuous
Concentrator not
operating, Power
Switch On.
operating, Power
Switch On.
Beeeeeeeep....
System Failure:
1. Unit overheating due
to blocked air intake.
1. Unit overheating due
to blocked air intake.
2. Insufficient power at
outlet.
outlet.
3. Internal repairs
required.
required.
1a. Remove and clean cabinet
filters.
filters.
1b. Move oxygen
concentrator at least three
inches away from walls,
draperies or furniture.
concentrator at least three
inches away from walls,
draperies or furniture.
2. DO NOT use extension
cords. Move to another
electrical outlet or circuit.
cords. Move to another
electrical outlet or circuit.
3. Call Service Provider.