SECTION 24—ALARM TEST
81
4. Replace tubing and tie wrap.
Time-Out Test
NOTE: For this procedure, refer to FIGURE 24.1.
1. With unit off ( ), disconnect the compressor connector from the main wiring harness.
2. Plug unit in and turn power on ( ).
3. The Time‐Out Failure alarm should activate within 40 seconds. Refer to SensO
Threshold on page 83 for shutdown mode. 
4. Replace tubing and tie wrap.
NOTE: If any alarm fails to perform to specification, contact Invacare Technical Service.
High Pressure Test
NOTE: For this procedure, refer to FIGURE 24.2 on page 82.
• High Pressure ‐ Occurs when pressure in product tank rises beyond a preset value, 
typically 23 p.s.i. ± 1 p.s.i. for 5 liter and 27 p.s.i. ± 1 p.s.i. for 10 liter.
• Main Valve Coil ‐ Alarm sequence occurs when the main valve coil, connection or 
P.C. board circuitry has failed.
1. With the unit running and flow set at 5 L/min., disconnect in‐line connector shown in 
2. The main valve coil alarm should activate within 40 seconds. Refer to SensO
3. Replace tubing and tie wrap.
NOTE: If any alarm fails to perform to specification, contact Invacare Technical Service.