Omega FOB100 Manuale Utente

Pagina di 30
 
15 
b) 
i=1.  In  case  of  a  probe  problem,  the  analog  output  for  this  channel  will  be  forced  to  its  maximum 
possible value (e.g., 10V or 20mA). 
c) 
i=2. In case of a probe problem, the analog output will be forced to its minimum possible value (e.g., 
0V or 4mA). 
 
“p[i]” This command allows the user to force the thermometer to enter in sleep mode (p+) or to exit it (p-). 
Once  in  sleep  mode,  the  thermometer  will  acquire  new  temperature  values  ONLY  when  the  “t”  or  “t[i]” 
command is received; when the t command is received, the thermometer will immediately scan all enabled 
channels, and the new temperature values will be displayed and sent through the RS-232 link, after which it 
will return again in sleep mode (there is no need to resend the p+ command again). 
a) 
When the thermometer is powered on, the sleep mode is always disabled. 
b) 
When  new  temperature  values  are  acquired,  these  new  values  will  be  shown  on  the  display  (if 
present) for about 40 seconds, after which “Sleep mode” will be displayed. 
c) 
The analog outputs will keep the last temperature readings, until new ones become available. 
d) 
The front panel led status lights will keep their values until new readings are acquired. 
 
“q” “q” allows for selecting the fast (q+) or the slow (q-) acquisition mode. This is equivalent to the Sampling 
Rate selection that is available on the FOB100 keypad. In slow mode, the last 4 temperatures are averaged 
before  being  displayed  or  sent  to  the  RS-232  port;  this  provides  a  better  reading  at  the  expense  of  being 
slower. 
 
 “s[j]”
 or “s[i  j]” This command will set the analog output span for the analog channel # i (if using a multi-
channel thermometer, the “i” parameter specifies the channel to which the command is addressed). The “j” 
parameter specifies the span that can range from 1.0 to 1000.0, in temperature units that are defined by the 
last “u” command. 
 
“t” or “t[i]” Returns the current temperature reading for channel # i. “---.-” is displayed if the specified channel 
number  is  not  currently  enabled  or  when  no  valid  temperature  is  available  for  the  requested  channel. 
However, during the initial warm-up time, the error “Err1” is sent in response to the “t” command. For multi-
channel instruments, if the “t” command is received without parameter, the temperatures for all channels are 
returned. All temperatures are returned in 
o
C or 
o
F unit, as specified by the “u” command. 
ex: 
“t4 
E
“ will return the temperature for channel # 4. 
 
“t 
E
”, with the FOB104 thermometer, will return: 
 
 
+24.3<CR> 
 
 
+24.5<CR> 
 
 
+24.7<CR> 
 
 
+24.4<CR> 
 
 
 
The  “t”  command  is  also  available  in  a  special  “ta”  version,  that  can  be  useful  for  continuously  and 
automatically  monitoring  all  enabled  channels.  This  special  mode  is  recommended  when  automatically 
acquiring  temperature  readings  with  a  suitable  data-logging  software,  and  can  be  activated  by  issuing  the 
“ta+”  command,  or  deactivated  with  the  “ta-“  command.  Once  activated,  the  following  information  is 
automatically sent when the temperature on a channel has been acquired: 
 
“C:I;T: ±xxx.x<CR>” 
where  i  is  the  channel  number,  and  xxx.x  is  the  temperature  reading  (signed  number).  For  the  FOB101 
model, i is always 1. 
 
 “u[i]” Set units of measurements to either Celsius (i=c) or Fahrenheit (i=f). Note that this unit selection is the 
same as the temperature unit selection made from the keypad (if the display option is installed). 
ex: 
“uc 
E
“, to select ºC (Celsius). 
 
“y” The “y” command can be used to confirm the “signal strength” of a probe and/or its connection. It can be 
used as a diagnostic tool to confirm the good operation of a probe and its patch cord (if applicable) connected 
to  a  channel.  The  returned  value  can  be  between  0.3  to  approximately  3.00;  a  higher  value  is  better.  A 
minimum of 1.25 is required for a channel to return a temperature.