Omega FOB100 Manuale Utente

Pagina di 30
 
16 
ex: 
“y 
E
”, with the FOB104 thermometer, will return: 
 
 
CH1: power:2.61, lamp attenuation: 250, CCD time: 50ms (fixed)<CR> 
 
 
CH2: power:2.51, lamp attenuation: 250, CCD time: 50ms (fixed)<CR> 
 
 
CH3: power:2.38, lamp attenuation: 125, CCD time: 50ms (fixed)<CR> 
 
 
CH4: power:1.09*, lamp attenuation: 125, CCD time: 50ms (fixed)<CR> 
 
*: 0.3 to 1.1 indicates that no probe is detected for this channel. 
 
 “z[j[“ or “z[i;j]” This command is used in conjunction with the “s” command, as it sets the temperature that 
corresponds  to  the  analog  zero  (minimum  value  of  analog  output).  For  multi-channel  models,  the  “i” 
parameter specifies the channel to which the command is addressed; valid values are 1, 2, 3 and 4. The “j” 
parameter  specifies  the  zero  temperature  value,  and  can  range  from  -100.0  to  1000.0,  in  units  that  are 
defined by the last “u” command. Factory default is -100.0. 
5.4  Typical temperature reading sequence 
Once  all  parameters  have  been  set,  the  following  sequence  should  normally  be  followed  to  extract 
temperature  information  from  your  FOB100  thermometer.  2  procedures  can  be  used  for  this  purpose.  The 
first method consists in using the “ta” command (preferred), while the second sequence would be: 
d) 
The host computer should send the “t” (or “t[i]”) command, followed by a 
E
 character. 
e) 
The  thermometer  responds  by  echoing  the  requested  temperature  values,  as  indicated 
above; the transmission is terminated by a “*” character. 
f) 
When  the  “*”  character  is  received,  it  is  suggested  to  immediately  terminate  the  dialog 
session by having the host computer issue the “r” command followed by a 
E
 character.