Omega OSP SERIES Manuale Utente

Pagina di 60
47
IBM 

IBM - PC
+ 5 V 
10 µF, 16 V 
10 µF, 16 V 
10 µF, 16 V 
15
Rx  
Tx
Ground
IBM 
IBM 
Tx
Rx
2 16 10
11
12
14
13
10 µF, 16 V 
10 µF, 16 V 
ICL 232 
TSC 232 
Tx
Rx
+ 5 V 
Mini-Din
TTL to RS 232 converter 
15.3
Communication protocol
The exchange of information when a OSP is interconnected with a PC is as it follows.
Serial Interface of PC has to be set to 9600 Baud, 8 bit and No parity.
15.3.1
Computer data request from OSP to PC
Computer
OSP 
Tx IDNAME
Rx IDNAME
Proceed if the name is acknowledged
Rx IDNAME
Tx IDNAME
If not, do not answer
Tx Instruction
Rx Instruction
Rx Instruction
Tx Instruction
Tx char
Rx char
Rx DATA 1
Tx DATA 1
Tx char
Rx char
Rx DATA 2
Tx DATA 2
Tx char
Rx char
Rx DATA 3
Tx DATA 3
Tx char
Rx char
Rx DATA 4
Tx DATA 4
Tx CHKSUM
Rx char
Rx CHKSUM
Tx CHKSUM
IDNAME, Instruction, DATA 1, DATA 2, DATA 3, DATA 4 and CHKSUM are 8-bit values (1 byte)
Tx CHKSUM = DATA 1+DATA 2+DATA 3+DATA 4 .AND. FF
Rx CHKSUM = DATA 1+DATA 2+DATA 3+DATA 4 .AND. FF
The above is useful to verify correct received data.
The minimum time-out of OSP  is 3 seconds.
IDNAME is set to 1.