Omega OSP SERIES Manuale Utente

Pagina di 60
48
Reading values
$00 - Pyrometer measure -------------------------------------------------------------- RX
DATA1
STATUS
DATA2
-
DATA3
VALUE (H)
DATA4
VALUE (L)
IF STATUS=0
MEASURE = SIGNED INTEGER ( VALUE ) / 10
IF STATUS=2
MEASURE = ERROR RJ
IF STATUS=10
MEASURE = OFL
IF STATUS=11
MEASURE = UFL
$01 - Emissivity ---------------------------------------------------------------------------- RX
DATA1
-
DATA2
-
DATA3
-
DATA4
VALUE
EMISSIVITY = VALUE / 100
$03 - Max input value -------------------------------------------------------------------- RX
DATA1
-
DATA2
-
DATA3
VALUE (H)
DATA4
VALUE (L)
MAX = SIGNED INTEGER ( VALUE ) / 10
$04 - Min input value--------------------------------------------------------------------- RX
DATA1
-
DATA2
-
DATA3
VALUE (H)
DATA4
VALUE (L)
MIN = SIGNED INTEGER ( VALUE ) / 10
$05 - Average input value -------------------------------------------------------------- RX
DATA1
-
DATA2
-
DATA3
VALUE (H)
DATA4
VALUE (L)
AVERAGE = SIGNED INTEGER ( VALUE ) / 10
$06 - Tag ID,Eng.unit,Tc type,IPTS ------------------------------------------------- RX
DATA1
Tag ID (0=Tag0, 1=TagA, 2=TagB,.....,19=TagS)
DATA2
Eng.unit (1=°C - 2=K - 3=°F)
DATA3
Tc type (0=Tc K - 1=Tc S)
DATA4
IPTS (0=68 - 1=90)
$07 - Tc measure -------------------------------------------------------------------------- RX
DATA1
STATUS
DATA2
-
DATA3
VALUE (H)
DATA4
VALUE (L)
IF STATUS=0
MEASURE = SIGNED INTEGER ( VALUE ) / 10
IF STATUS=1
MEASURE = ERROR RJ
IF STATUS=2
MEASURE = TC OFF
IF STATUS=10
MEASURE = OPTC
$08 - HAL Setpoint------------------------------------------------------------------------ RX
DATA1
-
DATA2
-
DATA3
VALUE (H)
DATA4
VALUE (L)
HAL = SIGNED INTEGER ( VALUE )
IF HAL=3001 THEN HAL=OFF
$09 - LAL Setpoint ------------------------------------------------------------------------ RX
DATA1
-
DATA2
-
DATA3
VALUE (H)
DATA4
VALUE (L)
LAL = SIGNED INTEGER ( VALUE )
IF LAL=3001 THEN LAL=OFF
$0B - First 4 chars tag description-------------------------------------------------- RX
DATA1
'X'  (ASCII CODE)
DATA2
'X'  (ASCII CODE)
DATA3
'X'  (ASCII CODE)