RayTek FA Manuale Utente

Pagina di 67
Installation 
20 
Marathon Series FA/FR 
 
 
Figure 15: Electronics Housing Wiring 
 
Incorrect wiring can damage the sensor and void the warranty! Before applying power, 
make sure all connections are correct and secure!
 
 
The following figure illustrates how to remove the terminal block. 
 
Figure 16: Removing the Terminal Block 
The electronics box has two compression fittings to provide water sealing for the fiber optics cable and 
the electronics cable. The compression fitting must be tightened with a wrench around the cables to 
achieve  water  sealing.  To  achieve  sealing  for  the  fiber  optics  cable,  hand  tighten  the  compression 
fitting around the cable, and then use a wrench to tighten another 1 1/2 to 2 turns. If the cable is too 
thin then it may be necessary to add a bushing or heat shrink material to increase the cable diameter in 
order to ensure sealing. Hand tighten the compression fitting around the cable, and then use a wrench 
to tighten another 1½ to 2 turns. 
4- wire cable 
12-wire cable 
Cable compression 
fitting or conduit adapter 
Cable compression 
fitting or conduit adapter 
+mA (green) 
-mA (brown) 
NO/NC (blue) 
COM (orange) 
TRIG (yellow) 
SHIELD (bare) 
+mA (green) 
-mA (white) 
+24V (red)
GND (black)
TxA (purple)
TxB (grey)
RxA (black)
RxB (white)
+24V (red)
GND (black)
twisted cable pairs 
Attention: Do not confuse twisted 
pair black wire with single black 
power wire