SETUP Menu 
Teledyne API - T100 UV Fluorescence SO2 Analyzer  
98 
5.4.5. DILUTION RATIO (OPTION) 
The dilution ratio is a software option that allows the user to compensate for any dilution 
of the sample gas before it enters the sample inlet. Once the degree of dilution is known, 
add an appropriate scaling factor to the analyzer’s SO2 concentration calculation so that 
the measurement range and concentration values reflect the undiluted values when 
shown on the instrument’s front panel display screen and reported via the analog and 
serial outputs. 
Using the dilution ratio option is a 4-step process: 
1.  Select reporting range units: Follow the procedure in Section 5.4.4 
2.  Select the range: Use the procedures in Section 5.4. Ensure that the SPAN value 
entered is the maximum expected concentration of the undiluted calibration gas and 
that the span gas is either supplied through the same dilution inlet system as the 
sample gas or has an appropriately lower actual concentration. For example, with a 
dilution set to 100, a 1 ppm gas can be used to calibrate a 100 ppm sample gas if 
the span gas is not routed through the dilution system. On the other hand, if a 100 
ppm span gas is used, it needs to pass through the same dilution steps as the 
sample gas. 
3.  Set the dilution factor as a gain (e.g., a value of 20 means 20 parts diluent and 1 
part of sample gas): 
 
SAMPLE          RANGE = 500.000 PPB        SO2 =XXX.X     
 
 < TST  TST >  CAL  
 
 
SETUP
 
 
 
 
SETUP 
 
      
PRIMARY SETUP MENU
 
  
 
 CFG  DAS
  RNGE  
PASS  CLK  MORE  
EXIT
 
 
 
 
SETUP    
RANGE CONTROL MENU
 
 
 
 MODE
 
 
 
 
SET  UNIT
  DIL
  
 
EXIT
 
DIL
 only appears 
if the dilution ratio 
option has been 
installed 
 
 
 
SETUP  
DIL FACTOR: 1.0 GAIN
 
 
 
 
 
 
 
 0         0          0        1          .0 
             
ENTR
  
EXIT
 
Toggle to set the dilution factor. 
 
This is the number by which the 
analyzer will multiply the SO
2
 
concentrations of the gas passing 
through the reaction cell. 
 
 
 
SETUP    
DIL FACTOR: 20.0 GAIN
 
 
 
 
 
 
 
 0          0          2         0         .0 
             
ENTR
 
 EXIT
 
EXIT
 ignores the 
new setting. 
 
ENTR 
accepts the 
new setting. 
SAMPLE          
ENTER SETUP PASS : 818
 
 
        
 8        1        8 
 
                      
ENTR
  
 EXIT
 
 
Figure 5-8: SETUP RNGE – Dilution Ratio 
06807C DCN6650