SETUP Menu 
Teledyne API - T100 UV Fluorescence SO2 Analyzer  
96 
5.4.4. RANGE UNITS 
The T100 can display concentrations in parts per billion (10
9
 mols per mol, PPB), parts 
per million (10
6
 mols per mol, PPM), micrograms per cubic meter (µg/m
3
, UGM) or 
milligrams per cubic meter (mg/m
3
, MGM). Changing units affects all of the display, 
analog outputs, COM port and DAS values for all reporting ranges regardless of the 
analyzer’s range mode.  
To change the concentration units: 
 
 
 
 
SETUP X.X       
CONC UNITS: PPM
 
PPM   PPB
   
UGM
   
MGM  
    ENTER   EXIT
 
SAMPLE          RANGE = 500.000 PPB        SO2 =XXX.X    
 
 < TST  TST >  CAL  
 
 
SETUP
 
 
 
 
SETUP X.X 
 
      
PRIMARY SETUP MENU
 
  
 
 CFG  DAS
  RNGE  
PASS  CLK  MORE  
EXIT
 
 
 
 
SETUP X.X 
RANGE CONTROL MENU
 
 
 
 
 
MODE
 
 
 
 
SET  
UNIT 
 
 
 
EXIT
 
 
 
 
SETUP X.X          
CONC UNITS: PPB
 
PPM
   PPB
   
UGM
   
MGM    
 
 
   
 ENTER
   EXIT
 
SAMPLE          
ENTER SETUP PASS : 818
 
 
        
 8        1        8 
 
                      
ENTR
  
 EXIT
 
Select the preferred 
concentration unit. 
EXIT returns 
to the main menu.  
ENTR
 accepts 
the new unit, 
EXIT
 returns 
to the SETUP 
menu. 
 
Figure 5-7: SETUP RNGE – Concentration Units Selection 
 
I
MPORTANT
 
I
MPACT ON 
R
EADINGS OR 
D
ATA
 
Concentrations displayed in mg/m
3
 and µg/m
3
 use 0°C and 760 Torr as 
standard temperature and pressure (STP). Consult your local regulations 
for the STP used by your agency. See Section 5.4.4.1 for co
nverting 
volumetric to mass units 
06807C DCN6650