Teledyne API - T100 UV Fluorescence SO2 Analyzer  
106 
5.8. SETUP – VARS: VARIABLES SETUP AND DEFINITION 
Through the SETUP>MORE>VARS menu there are several-user adjustable software 
variables that define certain operational parameters. Usually, these variables are 
automatically set by the instrument’s firmware, but can be manually re-defined using the 
VARS menu. Table 5-2 lists all variables that are available within the 818 password 
protected level. 
Table 5-2: Variable Names (VARS) Revision 1.0.3 
NO. 
VARIABLE 
DESCRIPTION ALLOWED 
VALUES
DAS_HOLD_OFF 
Changes the internal data acquisition system (DAS) hold-off time, 
which is the duration when data are not stored in the DAS because 
the software considers the data to be questionable. That is the 
case during warm-up or just after the instrument returns from one 
of its calibration modes to SAMPLE mode. DAS_HOLD_OFF can 
be disabled entirely in each DAS channel. 
Can be between 0.5 
and 20 minutes 
Default=15 min. 
TPC_ENABLE 
Enables or disables the temperature and pressure compensation 
(TPC) feature (refer to Section 13.7.3). 
ON/OFF 
RCELL_SET 
Sets the sample chamber temperature. Increasing or decreasing 
this temperature will increase or decrease the rate at which SO
2
decays into SO
2
 (refer to Section 13.1.1). 
Do not adjust this setting unless under the direction of Teledyne 
API Technical Support personnel. 
30º C - 70º C 
Default= 50º C 
IZS_SET 
Sets the IZS option temperature. Increasing or decreasing this 
temperature will increase or decrease the permeation rate of the 
IZS source (refer to Sections 3.3.2.4, 9.5, 9.6). 
30º C - 70º C 
Default= 50º C 
DYN_ZERO 
Dynamic zero automatically adjusts offset and slope of the SO
2
 
response when performing a zero point calibration during an 
AutoCal (refer to Section 9). 
ON/OFF 
DYN_SPAN 
Dynamic span automatically adjusts slope and slope of the SO
2
 
response when performing a zero point calibration during an 
AutoCal (refer to Section 9). 
Note that the DYN_ZERO and DYN_SPAN features are not 
allowed for applications requiring EPA equivalency. 
ON/OFF 
CONC_PRECISION 
Allows the user to set the number of significant digits to the right of 
the decimal point display of concentration and stability values. 
AUTO, 1, 2, 3, 4 
Default=AUTO 
CLOCK_ADJ 
Adjusts the speed of the analyzer’s clock. Choose the + sign if the 
clock is too slow, choose the - sign if the clock is too fast. 
-60 to +60 s/day 
 
SERVICE_CLEAR 
ON resets the service interval timer. Returns to OFF upon reset. 
ON/OFF 
TIME_SINCE_SVC 
Number of hours since last service. 
0-500000 
10 
SVC_INTERVAL 
Sets the intval between service reminders. 
0-100000 
 
I
MPORTANT
 
I
MPACT ON 
R
EADINGS OR 
D
ATA
 
There are more VARS available when using the password, 929, for 
configuration. Use caution when pressing any buttons while in this setup. 
Any changes made may alter the performance of the instrument or cause 
the instrument to not function properly. Note that if there is an accidental 
change to a setup parameter, press EXIT to discard the changes. 