SETUP Menu 
Teledyne API - T100 UV Fluorescence SO2 Analyzer  
108 
5.9. SETUP – DIAG: DIAGNOSTICS FUNCTIONS 
The SETUP>MORE>DIAG menu provides a series of diagnostic functions whose 
parameters are dependent on firmware revision (refer to Menu Tree, A-5, in Appendix 
A).  Table 5-3 describes the functions and provides a cross-reference to the details for 
each in the remainder of this section. These functions can be used as tools in a variety of 
troubleshooting and diagnostic procedures. 
Table 5-3:  T100 Diagnostic (DIAG) Functions 
DIAGNOSTIC FUNCTION AND MEANING 
FRONT 
PANEL 
MODE 
INDICATOR 
SECTION 
SIGNAL I/O:
 Allows observation of all digital and analog 
signals in the instrument. Allows certain digital signals such as 
valves and heaters to be toggled ON and OFF. 
DIAG I/O 
ANALOG OUTPUT:
 When entered, the analyzer performs an 
analog output step test. This can be used to calibrate a chart 
recorder or to test the analog output accuracy. 
DIAG AOUT 
ANALOG I/O CONFIGURATION:
 Analog input/output 
parameters are available for viewing and configuration. 
DIAG AIO 
OPTIC TEST:
 When activated, the analyzer performs an optic 
test, which turns on an LED located inside the sensor module 
near the PMT (Fig. 10-15). This diagnostic tests the response 
of the PMT without having to supply span gas. 
DIAG OPTIC 
ELECTRICAL TEST:
 When activated, the analyzer performs 
an electric test, which generates a current intended to simulate 
the PMT output to verify the signal handling and conditioning of 
the PMT preamp board. 
DIAG ELEC 
LAMP CALIBRATION:
 The analyzer records the current 
voltage output of the UV source reference detector.  This value 
is used by the CPU to calculate the lamp ration used in 
determining the SO
2
 concentration  
DIAG LAMP 
PRESSURE CALIBRATION:
 The analyzer records the current 
output of the sample gas pressure sensor.  This value is used 
by the CPU to compensate the SO
2
 concentration when the 
TPC feature is enabled. 
DIAG PCAL 
FLOW CALIBRATION:
 This function is used to calibrate the 
gas flow output signals of sample gas and ozone supply. 
These settings are retained when exiting DIAG. 
DIAG FCAL 
TEST CHAN OUTPUT:
 Configures the A4 analog output 
channel. 
DIAG TCHN 
 