Teledyne API - T100 UV Fluorescence SO2 Analyzer  
112 
5.9.3. ANALOG I/O CONFIGURATION 
Table 6-8 lists the analog I/O functions that are available in the T100. 
Table 5-4: DIAG - Analog I/O Functions 
SUB MENU 
FUNCTION 
AOUTS CALIBRATED: 
Shows the status of the analog output calibration (YES/NO) and initiates a calibration of all 
analog output channels.  
CONC_OUT_1 
Sets the basic electronic configuration of the A1 analog output (SO
2
). There are three options: 
 
RANGE:
 Selects the signal type (voltage or current loop) and full scale level of the 
output. 
 
REC_OFS:
 Allows setting a voltage offset (not available when RANGE is set to Current 
Loop (CURR). 
 
AUTO_CAL:
 Performs the same calibration as AOUT CALIBRATED, but on this one 
channel only. 
NOTE: Any change to RANGE or REC_OFS requires recalibration of this output. 
CONC_OUT_2 
Same as for CONC_OUT_1 but for analog channel 2 (SO
2
TEST OUTPUT 
Same as for CONC_OUT_1 but for analog channel 3 (TEST) 
CONC_OUT_3 
(Not available in the analyzer’s standard configuration; applies when optional sensor installed). 
AIN CALIBRATED 
Shows the calibration status (YES/NO) and initiates a calibration of the analog input channels. 
XIN1 


XIN8 
For each of 8 external analog inputs channels, shows the gain, offset, engineering units, and 
whether the channel is to show up as a Test function. 
 
 
Table 5-5:  Analog Output Voltage Ranges 
RANGE 
MINIMUM OUTPUT 
MAXIMUM OUTPUT 
0-0.1 V 
-5 mV 
+105 mV 
0-1 V 
-0.05 V 
+1.05 V 
0-5 V 
-0.25 V 
+5.25 V 
0-10 V 
-0.5 V 
+10.5 V 
The default offset for all ranges is 0 VDC. 
 