SETUP Menu 
Teledyne API - T100 UV Fluorescence SO2 Analyzer  
110 
5.9.1. SIGNAL I/O 
The signal I/O diagnostic mode allows a user to review and change the digital and 
analog input/output functions of the analyzer.  Refer to Appendix A-4 for a complete list 
of the parameters available for review under this menu. 
 
I
MPORTANT
 
I
MPACT ON 
R
EADINGS OR 
D
ATA
 
Any changes of signal I/O settings will remain in effect only until the 
signal I/O menu is exited. Exceptions are the ozone generator override 
and the flow sensor calibration, which remain as entered when exiting. 
Access the signal I/O test mode from the DIAG Menu (refer to Figure 5-16), then press:  
 
EXAMPLE 
 
 
 
DIAG  
SIGNAL I / O 
 
 PREV NEXT 
JUMP 
 
                       ENTR
 
  
EXIT
 
 
 
 
DIAG  I / O          
0) EXT_ZERO_CAL=OFF
 
 
PREV  NEXT
 JUMP 
                      
PRNT  EXIT
 
Press JUMP to go 
directly to a specific 
signal  
See Appendix A-4 for 
a complete list of 
available SIGNALS 
 
 
 
DIAG  I / O 
JUMP TO: 12
 
 
 
 1     2                       
 
        ENTR
  EXIT
 
EXAMPLE:  
Enter 12 to Jump to 
12)
 ST_SYSTEM_OK=ON 
 
 
 
DIAG  I / O        
12) 
 
ST_SYSTEM_OK = ON
 
 
  PREV NEXT
 
JUMP
 
                ON   
PRNT 
EXIT
 
Exit to return 
 to the  
DIAG
 menu 
Press NEXT & PREV to 
move between signal 
types. 
Pressing the PRNT button will send a formatted 
printout to the serial port and can be captured 
with a computer or other output device. 
Toggle ON/(OFF) button to 
change status. 
 
Figure 5-17: 
DIAG – Signal I/O Menu 