Teledyne API - T100 UV Fluorescence SO2 Analyzer  
Communications Setup and Operation 
 
 
143 
6.5.3. USB PORT FOR REMOTE ACCESS 
The analyzer can be operated through a personal computer by downloading the TAPI 
USB driver and directly connecting their respective USB ports. 
1.  Install the Teledyne T-Series USB driver on your computer, downloadable from the 
Teledyne API website under Help Center>Software Downloads (www.teledyne-
api.com/software). 
2.  Run the installer file: “TAPIVCPInstaller.exe” 
 
3.  Connect the USB cable between the USB ports on your personal computer and your 
analyzer.  The USB cable should be a Type A – Type B cable, commonly used as a 
USB printer cable. 
4.  Determine the Windows XP Com Port number that was automatically assigned to 
the USB connection.  (Start → Control Panel → System → Hardware → Device 
Manager).  This is the com port that should be set in the communications software, 
such as APIcom or Hyperterminal. 
 
Refer to the Quick Start (Direct Cable Connection) section of the Teledyne APIcom 
Manual, PN 07463. 
5.  In the instrument’s SETUP>MORE>COMM>COM2 menu, make the following settings: 