Teledyne API – T100 UV Fluorescence SO2 Analyzer 
Instrument Maintenance 
225 
11.1. MAINTENANCE SCHEDULE 
Table 11-1 is the recommended maintenance schedule for the T100. Please note that in certain environments with high levels of 
dust, humidity or pollutant levels some maintenance procedures may need to be performed more often than shown. 
Table 11-1: 
T100 Preventive Maintenance Schedule 
ITEM ACTION 
FREQUENCY 
CAL 
CHECK 
MANUAL 
SECTION 
DATE PERFORMED 
1
Particulate filter 
Change particle 
filter 
Weekly
 
No 
 
 
 
 
 
 
 
Verify test functions 
Review and 
evaluate 
Weekly No 
11.2
Appendix C  
 
 
 
 
 
 
 
Zero/span check 
Evaluate offset 
and slope 
Weekly -- 
 
 
 
 
 
 
 
1
Zero/span calibration 
Zero and span 
calibration 
Every 3 months
 
--  
 
 
 
 
 
 
 
1
External zero air 
scrubber (optional) 
Exchange 
chemical 
Every 3 months
 
No 
 
 
 
 
 
 
 
1
Perform flow check 
Check Flow 
Every 6 Months 
No 
 
 
 
 
 
 
 
Internal IZS 
Permeation Tube 
 Replace 
Annually 
YES 
 
 
 
 
 
 
 
Perform pneumatic 
leak check 
Verify Leak 
Tight 
Annually or after repairs 
involving pneumatics 
Yes 
 
 
 
 
 
 
 
2
Pump diaphragm 
Replace 
Annually 
Yes 
 Refer to 
diaphragm kit 
instructions 
 
 
 
 
 
 
 
 Calibrate UV Lamp 
Output 
Perform LAMP 
CAL
  
Prior to zero/span 
calibration or PMT 
hardware calibration 
--  
 
 
 
 
 
 
 
3
PMT sensor 
hardware calibration 
Low-level 
hardware 
calibration 
On PMT/ preamp 
changes  if  
0.7 < SLOPE or 
SLOPE >
1.3 
Yes 
 
 
 
 
 
 
 
1
Sample chamber 
optics  
Clean 
chamber, 
windows and 
filters 
As necessary
 
Yes 
 
 
 
 
 
 
 
1
Critical flow orifice & 
sintered filters 
Replace As 
necessary
 
Yes 
 
 
 
 
 
 
 
1
 These Items are required to maintain full warranty; all other items are strongly recommended. 
2
 A pump rebuild kit is available from Teledyne API’s Technical Support including all instructions and required parts (refer to Appendix B for part numbers). 
Replace desiccant bags each time the inspection plate for the sensor assembly is removed.  
06807C DCN6650