Teledyne API – T100 UV Fluorescence SO2 Analyzer 
Instrument Maintenance 
 
227 
11.2. PREDICTIVE DIAGNOSTICS 
The analyzer’s test functions can be used to predict failures by looking at trends in their 
values (refer to Table 11-2) and by comparing them values recorded for them at the 
factory and recorded on the T100 Final Test and Validation Data Form (Teledyne API 
P/N 04551) that was shipped with your analyzer.   
A convenient way to record and track changes to these parameters is the internal data 
acquisition system (DAS). Also, APICOM control software can be used to download 
and record these data for review even from remote locations (Section 7.3 discusses 
APICOM). 
Table 11-2: 
Predictive Uses for Test Functions 
BEHAVIOR 
TEST FUNCTION 
DAS 
FUNCTION 
CONDITION 
EXPECTED ACTUAL 
INTERPRETATION 
Fluctuating 
 Developing leak in pneumatic system 
Slowly 
increasing 
 Flow path is clogging up.  
- Check critical flow orifice & sintered filter.
- Replace particulate filter 
PRES SMPPRS 
sample gas 
Constant within 
atmospheric 
changes 
Slowly 
decreasing 
 Developing leak in pneumatic system to 
vacuum (developing valve failure) 
DRK PMT 
DRKPMT 
PMT output 
when UV Lamp 
shutter closed 
Constant within 
±20 of check-
out value 
Significantly 
increasing 
 PMT cooler failure 
 Shutter Failure 
 Change in instrument response 
At span with 
IZS option 
installed  
Constant 
response from 
day to day 
Decreasing 
over time 
 Degradation of IZS permeation tube 
SO
2
 
Concentration 
CONC1 
Standard 
configuration at 
span 
stable for 
constant 
concentration 
Decreasing 
over time 
 Drift of instrument response; UV Lamp 
output is excessively low. 
Slowly 
Decreasing 
 Flow path is clogging up.  
- Check critical flow orifice & sintered filter.
- Replace particulate filter 
SAMP FL 
SMPFLW 
Standard 
Operation 
Stable 
Fluctuating 
 Leak in gas flow path. 
Fluctuating or 
Slowly 
increasing 
 UV detector wearing out 
 UV source Filter developing pin holes 
LAMP RATIO 
LAMPR 
Standard 
Operation 
Stable and near 
100% 
Slowly 
decreasing 
 UV detector wearing out 
 Opaque oxides building up on UV source 
Filter  
 UV lamp aging 
06807C DCN6650