Maintenance 
 
Instruction 3015-4342 
43
 
5.3 Flashing Memory 
As updates to the instrument’s firmware become available, the instrument can be upgraded by flashing its 
memory as described below. 
 
IMPORTANT!
  After flashing memory, all operating parameters and sensor calibrations will have been 
erased. The operator must have access to the Factory menu in order to set up the following parameters 
before placing the instrument back into operation: 
- DIGIPOT 
- Leak rate of internal leak source 
- Leak gas of internal leak source 
- Leak temperature of internal leak source 
- Leak temperature coefficient of internal leak source 
- Pressure sensor calibration 
- Flow rate sensor calibration 
- Box temperature sensor calibration 
 
Items Required: 
•  9-pin-Female to 9-pin-Male Serial Cable (P/N 104-4027) 
•  Flash Utility Software and Latest Firmware Image File  
(downloaded from Website http://www.bacharach-inc.com/downloads.htm) 
•  Personal computer: 
o
 
Running Windows
®
 95/98/2000/NT/XP 
o
 
Open COM port 
 
Procedure: 
1.  Connect one end of a 9-pin female to 9-pin male serial cable to the instrument’s rear-panel 
RS-232 connector. 
 
2.  Connect the other end of the cable to and an open COM port connector on the computer (it may be 
necessary to use a 9-pin to 25-pin adapter to make the computer connection). 
 
NOTE:
 COM1 is the default port used by the Flash Memory Program. If COM1 is already in use by 
another device, then connect the instrument to the next available COM port. The software will need 
to be reconfigured as described in Step 9 to use this port. 
 
3.  With the instrument turned OFF, set DIP switches 1, 2, and 3 located on the instrument’s rear 
panel to their UP positions. 
 
 
 
 
 
 
4. 
Turn ON the instrument. Observe that the pump does not start and that the front panel display 
does not light. 
 
NOTE:
 The following steps assume that the operator is familiar with the Windows Operating System, 
and is knowledgeable in creating folders, copying files, and navigating the file system using Windows 
Explorer. If necessary, refer to the Windows help files for information on performing these operations. 
 
5.  Create a folder on the computer’s hard drive named “C:\RFU” (Rabbit Field Utility). 
 
6.  Copy the flash utility software and the latest firmware image file (xxx.bin – where “xxx” is the 
name and version number of the file) into the folder created in Step 5.