Setup Menu 
Teledyne API – Model T300/T300M CO Analyzer 
102 
5.4.3.1. SINGLE RANGE MODE (SNGL)  
 
Single Range Mode (SNGL) is the default reporting range mode for the analyzer. 
When the single range mode is selected (SNGL), all analog CO concentration outputs 
(A1 and A2) are slaved together and set to the same reporting range limits (e.g. 500.0 
ppb). The span limit of this reporting range can be set to any value within the physical 
range of the analyzer. 
Although both outputs share the same concentration reporting range, the electronic 
signal ranges of the analog outputs may still be configured for different values (e.g. 0-5 
VDC, 0-10 VDC, etc; see Section 5.9.3.1)  
To select SNGL range mode and to set the upper limit of the range, press: 
 
06864B DCN6314