Setup Menu 
Teledyne API – Model T300/T300M CO Analyzer 
120 
5.9.3. ANALOG I/O CONFIGURATION  
The T300/T300M Analyzer comes equipped with four analog outputs.   
  The first two outputs (A1 & A2) carry analog signals that represent the currently 
measured concentration of CO (see Section 5.4.1).   
  The third output (A3) is only active if the analyzer is equipped with one of the 
optional 2
nd
 gas sensors (e.g. O
2
 or CO
2
). 
  The fourth output (A4) outputs a signal that can be set to represent the current value 
of one of several test functions (see Table 5-9)
Table 5-5 lists the analog I/O functions that are available in the T300/T300M Analyzer. 
Table 5-5: 
DIAG - Analog I/O Functions 
SUB MENU 
OUTPUT 
CHANNEL 
FUNCTION 
AOUT 
CALIBRATED 
ALL 
Initiates a calibration of the A1A2, A3 and A4 analog output channels that 
determines the slope and offset inherent in the circuitry of each output.  
These values are stored and applied to the output signals by the CPU 
automatically. 
CONC_OUT_1
 
A1 
Sets the basic electronic configuration of the A1 output (CO Concentration).   
There are four options: 
  RANGE
1
:  Selects the signal type (voltage or current loop) and level of the 
output. 
  REC OFS
1
: Allows them input of a DC offset to let the user manually adjust 
the output level. 
  AUTO CAL: Enables / Disables the AOUT CALIBRATED feature. 
  CALIBRATED: Performs the same calibration as AOUT CALIBRATED
but on this one channel only.   
CONC_OUT_2
 
A2 
  Same as for CONC_OUT_1 but for analog channel A2
CONC_OUT_3
 
A3 
  Same as for CONC_OUT_1 but for analog channel A3 but only if either the 
optional O
2
 or CO
2
 sensors are installed. 
TEST OUTPUT
 
A4 
  Same as for CONC_OUT_1 but for analog channel A4 (TEST CHANNEL). 
AIN 
CALIBRATED 
N/A
 
Initiates a calibration of the A-to-D Converter circuit located on the 
Motherboard. 
XIN1 


XIN8 
 
For each of 8 external analog input channels, shows the gain, offset, 
engineering units, and whether the channel is to show up as a Test function. 
Any changes made to RANGE or REC_OFS require recalibration of this output. 
06864B DCN6314