Setup Menu 
Teledyne API – Model T300/T300M CO Analyzer 
128 
5.9.3.6. MANUAL CALIBRATION OF THE ANALOG OUTPUTS CONFIGURED FOR VOLTAGE 
RANGES 
For highest accuracy, the voltages of the analog outputs can be manually calibrated.   
Note 
The menu for manually adjusting the analog output signal level will only 
appear if the AUTO-CAL feature is turned off for the channel being 
adjusted (see Section 5.9.3.3). 
Calibration is performed with a voltmeter connected across the output terminals and by 
changing the actual output signal level using the front panel buttons in 100, 10 or 1 
count increments.  See Figure 3-9 for pin assignments and diagram of the analog output 
connector. 
+DC     Gnd    
  
Figure 5-4: 
Setup for Checking / Calibrating DCV Analog Output Signal Levels  
 
Table 5-7:  Voltage Tolerances for the TEST CHANNEL Calibration 
FULL SCALE 
ZERO TOLERANCE
SPAN VOLTAGE 
SPAN 
TOLERANCE 
MINIMUM ADJUSTMENT 
(1 count) 
0.1 VDC 
±0.0005V 
90 mV 
±0.001V 
0.02 mV 
1 VDC 
±0.001V 
900 mV 
±0.001V 
0.24 mV 
5 VDC 
±0.002V 
4500 mV 
±0.003V 
1.22 mV 
10 VDC 
±0.004V 
4500 mV 
±0.006V 
2.44 mV 
 