Teledyne t320u Manuale Utente

Pagina di 426
Teledyne API – Model T300/T300M CO Analyzer 
Setup Menu 
141 
5.10.1. SETTING THE T300 CONCENTRATION ALARM LIMITS 
To enable either of the CO concentration alarms and set the limit points, press: 
SETUP X.X         
CO ALRM 2: OFF
ON
ENTR EXIT
SETUP X.X       
 PRIMARY SETUP MENU
 CFG DAS RNGE PASS CLK
MORE
EXIT
SETUP X.X         
SECONDARY SETUP MENU
 COMM VARS DIAG
ALRM
EXIT
DIAG FCAL         CO ALARM 2=300.00 PPM
3
0
0
.0
0
ENTR
EXIT
Toggle these 
buttons to Set 
the alarm point
EXIT discards the new 
setting
ENTR accepts the 
new setting
SAMPLE
RANGE=50.0 PPM          CO= XX.XX
<TST
TST> CAL
SETUP
SETUP X.X         
CO ALRM 2, DISALBED
NEXT
EDIT
PRNT EXIT
Toggle this button to 
enable/disable 
the alarm
Continue pressing NEXT  until the desired 
Alarm is selected
 
06864B DCN6314