Teledyne API – Model T300/T300M CO Analyzer 
Communications Setup and Operation 
147 
6.2.3. COM PORT TESTING 
The serial ports can be tested for correct connection and output in the COMM menu. 
This test sends a string of 256 ‘w’ characters to the selected COM port. While the test is 
running, the red LED labeled TX for that COM port on the instrument’s rear panel 
analyzer should flicker.  
To initiate the test, access the COMMUNICATIONS Menu (SETUP>MORE>COMM), 
then press: 
 
 
Select which 
COMM port to 
test. 
 
 
 
SETUP X.X
COMMUNICATIONS MENU
 
  ID    INET    
 COM1    
 COM2
 
   
  
EXIT
 
 
 
 
SETUP X.X          
COM1 : TEST PORT
 
<SET           
   TEST   
      
 
EXIT
 
 
 
 
SETUP X.X         
TRANSMITTING TO COM1
 
<SET           
 
  TEST 
  
 
 
  
 
EXIT 
 
 
 
SETUP X.X
COM1 MODE:0
 
      
 SET>   
 EDIT
   
       
 
 
  EXIT
 
SETUP X.X          COM1 BAUD RATE:19200
 
 
<SET  
SET>   
 EDIT
   
 
   
 
EXIT
 
EXIT returns to 
COMM menu 
Test runs 
automatically 
 
Figure 6-3: 
COMM – COM1 Test Port 
6.3. RS-232 
The RS232 and COM2 communications (COMM) ports operate on the RS-232 protocol 
(default configuration). Possible configurations for these two COMM ports are 
summarized as follows:   
  RS232 port can also be configured to operate in single or RS-232 Multidrop mode 
(Option 62); refer to Sections 3.3.1.8 and 5.7.1. 
  COM2 port can be left in its default configuration for standard RS-232 operation 
including multidrop, or it can be reconfigured for half-duplex RS-485 operation 
(please contact the factory for this configuration).  
Note that when the rear panel COM2 port is in use, except for multidrop 
communication, the rear panel USB port cannot be used. (Alternatively, when the USB 
port is enabled, COM2 port cannot be used except for multidrop). 