Teledyne API – Model T300/T300M CO Analyzer 
Communications Setup and Operation 
165 
6.7.3.4. SETTING HESSEN PROTOCOL STATUS FLAGS 
Teledyne API’s implementation of Hessen protocols includes a set of status bits that the 
instrument includes in responses to inform the host computer of its condition.  Each bit 
can be assigned to one operational and warning message flag.  The default settings for 
these bit/flags are: 
Table 6-6:  Default Hessen Status Flag Assignments 
STATUS FLAG NAME 
DEFAULT BIT ASSIGNMENT 
WARNING FLAGS 
SAMPLE FLOW WARNING 
0001 
BENCH TEMP WARNING 
0002 
SOURCE WARNING
 
 
0004 
BOX TEMP WARNING
 
 
0008 
WHEEL TEMP WARNING 
0010 
SAMPLE TEMP WARN 
0020 
SAMPLE PRESS WARN 
0040
 
INVALID CONC 
 
(The Instrument’s Front Panel Display Will Show The 
Concentration As “Warnings”) 
0080
 
OPERATIONAL FLAGS
1
 
Instrument OFF 0100 
In MANUAL Calibration Mode 
0200 
In ZERO Calibration Mode
4
 0400
 
In 
O
2
 Calibration Mode (if 
O
2
 sensor installed )
2,4 
0400 
In 
CO
2
 Calibration Mode (if 
CO
2
 sensor installed )
2,4 
0400 
In SPAN Calibration Mode 
0800 
UNITS OF MEASURE FLAGS 
 
UGM 
0000 
MGM 
2000 
PPB 
4000 
PPM 
6000 
SPARE/UNUSED BITS 
1000, 8000 
UNASSIGNED FLAGS (0000) 
AZERO WARN
2
 
DCPS WARNING 
CANNOT DYN SPAN
2
 
REAR BOARD NOT DET 
CANNOT DYN ZERO
3
 SYNC 
WARNING
1
 
CONC ALARM 1
3
 
SYSTEM RESET
1
 
CONC ALARM 2
3
 
 
1
 
These status flags are standard for all instruments and should probably not be modified.
 
2
 
Only applicable if the optional internal span gas generator is installed. 
3
    Only applicable if the analyzer is equipped with an alarm options. 
3
 
It is possible to assign more than one flag to the same Hessen status bit.  This allows the grouping of 
similar flags, such as all temperature warnings, under the same status bit.   
 
Be careful not to assign conflicting flags to the same bit as each status bit will be triggered if any of the 
assigned flags is active.   