Teledyne API – Model T300/T300M CO Analyzer 
Data Acquisition System (DAS) and APICOM 
169 
7.1.1. DAS DATA CHANNELS 
The key to the flexibility of the DAS is its ability to store a large number of 
combinations of triggering events and data parameters in the form of data channels.  
Users may create up to 50 data channels and each channel can contain one or more 
parameters.  For each channel, the following are selected: 
  One triggering event is selected. 
  Up to 50 data parameters, which can be the shared between channels.   
  Several other properties that define the structure of the channel and allow the user 
to make operational decisions regarding the channel. 
Table 7-2:  DAS Data Channel Properties 
PROPERTY DESCRIPTION 
DEFAULT 
SETTING 
SETTING RANGE 
NAME 
The name of the data channel. 
“NONE” 
Up to 6 letters or digits 
1.  
TRIGGERING 
EVENT 
The event that triggers the data channel to measure 
and store the datum. 
ATIMER 
Any available event  
(see Appendix A-5). 
NUMBER AND 
LIST OF 
PARAMETERS 
A User-configurable list of data types to be 
recorded in any given channel. 
(COMEAS) 
Any available parameter 
(see Appendix A-5). 
REPORT PERIOD 
The amount of time between each channel data 
point. 
000:01:00 
(1 hour)
 
000:00:01 to 
366:23:59 
(Days:Hours:Minutes) 
NUMBER OF 
RECORDS 
The number of reports that will be stored in the data 
file.  Once the limit is exceeded, the oldest data is 
over-written. 
100 
1 to 1 million, limited by 
available storage space. 
RS-232 REPORT 
Enables the analyzer to automatically report 
channel values to the RS-232 ports. 
OFF 
OFF or ON 
CHANNEL 
ENABLED 
Enables or disables the channel.  Allows a channel 
to be temporarily turned off without deleting it. 
ON 
OFF or ON 
CAL HOLD OFF 
Disables sampling of data parameters while 
instrument is in calibration mode 
2
OFF 
OFF or ON 
1
 More with APICOM, but only the first six are displayed on the front panel. 
2
 When enabled records are not recorded until the DAS HOLD OFF period is passed after calibration mode.  DAS HOLD OFF SET in 
the VARS menu (see Section 7.1.11
7.1.2. DEFAULT DAS CHANNELS 
A set of default Data Channels has been included in the analyzer’s software for logging 
CO concentration and certain predictive diagnostic data.  These default channels include 
but are not limited to: 
  CONC: Samples CO concentration at one minute intervals and stores an average 
every hour with a time and date stamp.  Readings during calibration and calibration 
hold off are not included in the data.   
  By default, the last 800 hourly averages are stored. 
  PNUMTC: Collects sample flow and sample pressure data at five-minute intervals 
and stores an average once a day with a time and date stamp.  This data is useful 
for monitoring the condition of the pump and critical flow orifice (sample flow) and 
the sample filter (clogging indicated by a drop in sample pressure) over time to 
predict when maintenance will be required.   
  The last 360 daily averages (about 1 year) are stored. 