Data Acquisition System (DAS) and APICOM 
Teledyne API – Model T300/T300M CO Analyzer 
178 
7.1.7. SAMPLE PERIOD AND REPORT PERIOD 
The DAS defines two principal time periods by which sample readings are taken and 
permanently recorded: Sample and Report periods.  
  SAMPLE PERIOD:  Determines how often DAS temporarily records a sample 
reading of the parameter in volatile memory.  SAMPLE PERIOD is only used when 
the DAS parameter’s sample mode is set for AVGSDEVMIN or MAX. 
The SAMPLE PERIOD is set to one minute by default and generally cannot be 
accessed from the standard DAS front panel menu, but is available via the 
instrument’s communication ports by using APICOM or the analyzer’s standard 
serial data protocol.   
  REPORT PERIOD:  Sets how often the sample readings stored in volatile memory 
are processed, (e.g.  average, minimum or maximum are calculated) and the results 
stored permanently in the instrument’s Disk-on-Module as well as transmitted via the 
analyzer’s communication ports.  The Report Period may be set from the front 
panel.  If the INST sample mode is selected the instrument stores and reports an 
instantaneous reading of the selected parameter at the end of the chosen report 
period. 