Teledyne t320u Manuale Utente

Pagina di 426
Data Acquisition System (DAS) and APICOM 
Teledyne API – Model T300/T300M CO Analyzer 
184 
7.1.9.2. THE STARTING DATE FEATURE 
This option allows the user to specify a starting date for any given channel in case the 
user wants to start data acquisition only after a certain time and date.  If the STARTING 
DATE
 is in the past (the default condition), the DAS ignores this setting and begins 
recording data as defined by the REPORT PERIOD setting. 
The  STARTING DATE generally cannot be accessed from the standard DAS front 
panel menu, but is available via the instrument’s communication ports by using 
APICOM or the analyzer’s standard serial data protocol. 
7.1.10. DISABLING/ENABLING DATA CHANNELS 
Data channels can be temporarily disabled, which can reduce the read/write wear on the 
Disk-on-Module.   
To disable a data channel, follow the instruction shown in Section 7.1.4 then press: 
 
 
06864B DCN6314