Calibration Procedures 
205 
9.3. MANUAL CALIBRATION WITH ZERO/SPAN VALVES 
There are a variety of valve options available on the T300/T300M for handling 
calibration gases (see Table 1-1 for descriptions of each).   
Generally performing calibration checks and zero/span point calibrations on analyzers 
with these options installed is similar to the methods discussed in the previous sections 
of this section. The primary differences are: 
  On instruments with Z/S valve options, zero air and span gas is supplied to the 
analyzer through other gas inlets besides the sample gas inlet. 
  The zero and span calibration operations are initiated directly and independently 
with dedicated buttons (CALZ & CALS). 
 
9.3.1. SETUP FOR CALIBRATION USING VALVE OPTIONS 
Each of the various calibration valve options requires a different pneumatic setup that is 
dependent on the exact nature and number of valves present.   
Source of
SAMPLE GAS
Removed during 
calibration
MODEL 701
Zero Gas 
Generator
Calibrated 
CO Gas
at span gas 
concentration
VENT 
here if input 
is pressurized
PRESSURE SPAN
ZERO AIR
SAMPLE
EXHAUST
VENT SPAN
Model 700 gas 
Dilution 
Calibrator
VENT
VENT
Instrument 
Chassis
 
Figure 9-3: 
Pneumatic Connections – Option 50A: Ambient Zero/Ambient Span Calibration Valves 
 