12.1.3.  THE DIAGNOSTIC SIGNAL I/O FUNCTION 
The signal I/O diagnostic mode allows access to the digital and analog I/O in the 
analyzer.  Some of the digital signals can be controlled through the touchscreen.  These 
signals, combined with a thorough understanding of the instruments Theory of 
Operation (found in Section 13), are useful for troubleshooting in three ways: 
  The technician can view the raw, unprocessed signal level of the analyzer’s critical 
inputs and outputs.   
  Many of the components and functions that are normally under algorithmic control of 
the CPU can be manually exercised.   
  The technician can directly control the signal level Analog and Digital Output signals. 
This allows the technician to observe systematically the effect of directly controlling 
these signals on the operation of the analyzer.  The following flowchart shows an 
example of how to use the Signal I/O menu to view the raw voltage of an input signal or 
to control the state of an output voltage or control signal.  (See also Sections 5.9.1 and 
12.5.8.1). 
 
06864B DCN6314