Teledyne API – Model T300/T300M CO Analyzer 
Troubleshooting and Service 
265
 
12.1.4.3. RELAY BOARD STATUS LEDS 
There are eight LEDs located on the Relay Board.  The most important of which is D1, 
which indicates the health of the I
2
C bus.  If D1 is blinking the other faults following 
LEDs can be used in conjunction with DIAG menu signal I/O to identify hardware 
failures of the relays and switches on the relay (see Section 12.1.3  and Appendix D).   
Table 12-4:  I
2
C Status LED Failure Indications 
LED 
FUNCTION 
FAULT STATUS 
INDICATED FAILURE(S) 
D1 
(Red) 
I
2
C bus Health 
(Watch Dog 
Circuit) 
Continuously ON  
or 
Continuously OFF 
Failed/Halted CPU 
Faulty Motherboard, Touchscreen or Relay Board 
Faulty Connectors/Wiring between Motherboard, 
Touchscreen or Relay Board 
Failed/Faulty +5 VDC Power Supply (PS1) 
 
 
 
STATUS LED’s 
DC VOLTAGE TEST 
POINTS 
RELAY PCA 
PN 04135 
 
Figure 12-5: 
Relay Board Status LEDs 