Troubleshooting and Service 
279
 
12.5.2. DC POWER SUPPLY 
If you have determined that the analyzer’s AC mains power is working, but the unit is 
still not operating properly, there may be a problem with one of the instrument’s 
switching power supplies.  The supplies can have two faults, namely no DC output, and 
noisy output.   
To assist tracing DC Power Supply problems, the wiring used to connect the various 
printed circuit assemblies and DC Powered components and the associated test points on 
the relay board follow a standard color-coding scheme as defined in the following table.   
 
Table 12-6: 
DC Power Test Point and Wiring Color Codes 
NAME 
TEST POINT# 
TP AND WIRE COLOR 
Dgnd 1 
Black 
+5V 2 
Red 
Agnd 3 
Green 
+15V 4 
Blue 
-15V 5 
Yellow 
+12R 6 
Purple 
+12V 7 
Orange 
 
A voltmeter should be used to verify that the DC voltages are correct per the values in 
the table below, and an oscilloscope, in AC mode, with band limiting turned on, can be 
used to evaluate if the supplies are producing excessive noise (> 100 mV p-p).   
Table 12-7:  DC Power Supply Acceptable Levels 
CHECK RELAY BOARD TEST POINTS 
FROM TEST POINT 
TO TEST POINT 
POWER 
SUPPLY 
ASSY 
VOLTAGE
NAME # NAME # 
MIN V 
MAX V 
PS1 +5 Dgnd 
1 +5 2 
4.8 
5.25 
PS1 +15 Agnd 3  +15  4 
13.5 
16V 
PS1 -15 Agnd 3 -15V 5 
-14V 
-16V 
PS1 Agnd Agnd  3  Dgnd  1 -0.05 
0.05 
PS1 Chassis Dgnd  1  Chassis N/A -0.05 0.05 
PS2 +12 
+12V 
Ret 
6 +12V 7 
11.75 
12.5 
PS2 Dgnd 
+12V 
Ret 
6  Dgnd  1 -0.05 
0.05 
 
12.5.3. I
2
C BUS 
Operation of the I
2
C bus can be verified by observing the behavior of D1 on the relay 
PCA. Assuming that the DC power supplies are operating properly, the I
2
C bus is 
operating properly if D1 on the relay PCA is flashing. 
06864B DCN6314