Teledyne API – Model T300/T300M CO Analyzer 
Troubleshooting and Service 
283
 
12.5.7.6. PRESSURE/FLOW SENSOR ASSEMBLY 
The pressure/flow sensor PCA, located on the top of the absorption bench, can be 
checked with a voltmeter using the following procedure which, assumes that the wiring 
is intact, and that the motherboard and the power supplies are operating properly: 
1.  For Pressure related problems: 
  Measure the voltage across C1 - it should be 5 ± 0.25 VDC.  
 If not, then the board is bad. 
  Measure the voltage across TP4 and TP1.   
  With the sample pump disabled it should be 4500 mV ±250 mV.   
  With the pump energized it should be approximately 200 mV less.  
 If the voltage is incorrect, then S1, the pressure transducer is bad, the board is bad, or 
there is a pneumatic failure preventing the pressure transducer from sensing the 
absorption cell pressure properly. 
2.  For flow related problems: 
  Measure the voltage across TP2 and TP1 - it should be 10 ±0.25 VDC.  
 If not, then the board is bad. 
  Measure the voltage across TP3 and TP1:   
  With proper flow (800 sccm at the sample inlet) this should be approximately 
4.5V (this voltage will vary with altitude).   
  With flow stopped (sample inlet blocked) the voltage should be approximately 
1V.   
If the voltage is incorrect, the flow sensor is bad, the board is bad, or there is a leak 
upstream of the sensor. 