Troubleshooting and Service 
Teledyne API – Model T300/T300M CO Analyzer 
286 
12.5.8.4. STATUS OUTPUTS 
The procedure below can be used to test the Status outputs: 
1.  Connect a jumper between the “D“ pin and the “” pin on the status output 
connector. 
2.  Connect a 1000 ohm resistor between the “+” pin and the pin for the status output 
that is being tested. 
3.  Connect a voltmeter between the “” pin and the pin of the output being tested (see 
table below).   
Under the DIAG SIGNAL I/O menu (see Section 12.1.3), scroll through the inputs 
and outputs until you get to the output in question.  Alternately turn on and off the 
output noting the voltage on the voltmeter, it should vary between 0 volts for ON and 5 
volts for OFF.   
Table 12-12: 
Status Outputs Check 
PIN (LEFT TO RIGHT) 
STATUS 
1 SYSTEM 
OK 
2 CONC 
VALID 
3 HIGH 
RANGE 
4 ZERO 
CAL 
5 SPAN 
CAL 
6 DIAG 
MODE 
7 SPARE 
8 SPARE 
 
12.5.8.5. CONTROL INPUTS – REMOTE ZERO, SPAN 
The control input bits can be tested by the following procedure: 
1.  Jumper the +5 pin on the Status connector to the U on the Control In connector.   
2.  Connect a second jumper from the 
pin on the Status connector to the A pin on the 
Control In connector.  The instrument should switch from Sample Mode to ZERO 
CAL R mode.   
3.  Connect a second jumper from the 
pin on the Status connector to the B pin on the 
Control In connector.  The instrument should switch from Sample Mode to SPAN 
CAL R mode. 
4.  In each case, the T300/T300M should return to Sample Mode when the jumper is 
removed. 