Teledyne t320u Manuale Utente

Pagina di 426
Troubleshooting and Service 
Teledyne API – Model T300/T300M CO Analyzer 
290 
12.6.2. REMOVING/REPLACING THE GFC WHEEL  
When removing or replacing the GFC Wheel it is important to perform the disassembly 
in the following order to avoid damaging the components: 
1.  Turn off the analyzer. 
2.  Remove the top cover. 
3.  Open the instrument’s hinged front panel.   
4.  Locate the GFC Wheel/motor assembly.  See Figure 3-6. 
5.  Unplug the following electronic components: 
  The GFC Wheel housing temperature sensor 
 GFC 
Wheel 
heater 
  GFC Wheel motor power supply 
SAFETY SHIELD
SYNCHRONOUS MOTOR
SOURCE ASSEMBLY
THERMISTOR
HEATER
 
Figure 12-15:  Opening the GFC Wheel Housing 
 
6.  Remove the two (2) screws holding the opto-pickup printed circuit assembly to the 
GFC Wheel housing. 
06864B DCN6314