Theory of Operation 
Teledyne API – Model T300/T300M CO Analyzer 
314 
The four sample and hold circuits are designated as follows: 
Table 13-2: 
Sync DEMOD Sample and Hold Circuits 
Active When: 
Designation 
IR BEAM PASSING THROUGH 
Segment Sensor Pulse is: 
Measure Gate 
MEASUREMENT cell of GFC Wheel 
HIGH 
Measure Dark Gate 
MEASUREMENT Cell of GFC Wheel 
LOW 
Reference Gate 
REFERENCE cell of GFC Wheel 
HIGH 
Reference Dark Gate 
REFERENCE cell of GFC Wheel 
LOW 
Timing for activating the Sample and Hold Circuits is provided by a Phase Lock Loop 
(PLL) circuit.  Using the segment sensor output as a reference signal the PLL generates 
clock signal at ten times that frequency.  This faster clock signal is used by the PLD to 
make the Sample and Hold Circuits capture the signal during the center portions of the 
detected waveform, ignore the rising and falling edges of the detector signal. 
 
Sample & Hold 
Active 
Detector 
Output 
Sample & Hold 
Inactive 
 
Figure 13-14:  Sample & Hold Timing  
13.4.3.2. SYNC/DEMOD STATUS LEDS 
The following two status LEDs located on the sync/demod board provide additional 
diagnostic tools for checking the GFC Wheel rotation.   
Table 13-3: 
Sync/Demod Status LED Activity 
LED 
Function 
Status OK 
Fault Status 
D1 
M/R Sensor Status 
LED flashes approximately  
2/second 
LED is stuck 
ON or OFF 
D2 
Segment Sensor Status 
LED flashes approximately  
6/second 
LED is stuck 
ON or OFF 
See Section 12.1.4.2 for more information. 