Teledyne API – Model T300/T300M CO Analyzer 
Theory of Operation 
319
 
13.4.5.3. THERMISTOR INTERFACE 
This circuit provides excitation, termination and signal selection for several negative-
coefficient, thermistor temperature sensors located inside the analyzer.  They are as 
follows: 
S
AMPLE 
T
EMPERATURE 
S
ENSOR
 
The source of this signal is a thermistor located inside the sample chamber of the Optical 
Bench.  It measures the temperature of the sample gas in the chamber.  This data is used 
to during the calculation of the CO concentration value. 
B
ENCH 
T
EMPERATURE 
S
ENSOR
 
This thermistor is attached to the sample chamber housing. It reports the current 
temperature of the chamber housing to the CPU as part of the bench heater control loop. 
W
HEEL 
T
EMPERATURE 
S
ENSOR
 
This thermistor is attached to the heatsink on the GFC Wheel motor assembly. It reports 
the current temperature of the wheel/motor assembly to the CPU as part of the Wheel 
Heater control loop. 
B
OX 
T
EMPERATURE 
S
ENSOR
 
A thermistor is attached to the motherboard.  It measures the analyzer’s internal 
temperature.  This information is stored by the CPU and can be viewed by the user for 
troubleshooting purposes via the front panel display (see Section 12.1.2). 
13.4.5.4. ANALOG OUTPUTS 
The analyzer comes equipped with four analog outputs: A1A2A3 and A4.   The type 
of data and electronic performance of these outputs are configurable by the user (see 
Section 5.4). 
O
UTPUT 
L
OOP
-
BACK
 
All four analog outputs are connected back to the A/D converter through a loop-back 
circuit.  This permits the voltage outputs to be calibrated by the CPU without need for 
any additional tools or fixtures.   
13.4.5.5. INTERNAL DIGITAL I/O 
This channel is used to communicate digital status and control signals about the 
operation of key components of the Optical Bench.  The CPU sends signals to the 
sync/demod board that initiate the ELECTRICAL TEST and DARK CALIBRATION 
procedures.   
13.4.5.6. EXTERNAL DIGITAL I/O 
This External Digital I/O performs two functions: status outputs and control inputs. 
S
TATUS 
O
UTPUTS
 
Logic-Level voltages are output through an optically isolated 8-pin connector located on 
the rear panel of the analyzer.  These outputs convey good/bad and on/off information 